ABBOTT M2000RT 09K15-01

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-05-01 for ABBOTT M2000RT 09K15-01 manufactured by Abbott Molecular, Inc..

Event Text Entries

[21358439] A technical review by a product support engineer determined that the m 2000rt block drive assembly failure was caused by electrical arcing or a short circuit. The overheating was due to the thermal degradation of the electrical component and associated insulation which caused a small amount of blackening or soot marks.
Patient Sequence No: 1, Text Type: N, H10


[21459906] The abbott m2000rt is part of the m2000 realtime system and is for real-time pcr amplification and detection of nucleic acids. A customer in (b)(6), reported an error code display during initialization of the m2000rt instrument. The abbott field service engineer visited the customer site as a follow up to the customer report and observed a charred mark and some damaged components on the block drive assembly interface. The fse replaced the block drive assembly. There was no injury caused by this incident.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3005248192-2012-00015
MDR Report Key2840231
Report Source07
Date Received2012-05-01
Date of Report2012-04-02
Date of Event2012-03-27
Date Mfgr Received2012-04-02
Date Added to Maude2012-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJEAN LEETE
Manufacturer Street1300 EAST TOUHY AVE.
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal60018
Manufacturer Phone2123617274
Manufacturer G1ABBOTT MOLECULAR, INC.
Manufacturer Street1300 E. TOUHY AVE.
Manufacturer CityDES PLAINES IL 60018
Manufacturer CountryUS
Manufacturer Postal Code60018
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameABBOTT M2000RT
Generic NameREAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM
Product CodeOOI
Date Received2012-05-01
Catalog Number09K15-01
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MOLECULAR, INC.
Manufacturer Address1300 E. TOUHY AVE. DES PLAINES IL 60018 US 60018


Patients

Patient NumberTreatmentOutcomeDate
10 2012-05-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.