MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2012-05-01 for ABBOTT M2000RT 09K15-01 manufactured by Abbott Molecular, Inc..
[21358439]
A technical review by a product support engineer determined that the m 2000rt block drive assembly failure was caused by electrical arcing or a short circuit. The overheating was due to the thermal degradation of the electrical component and associated insulation which caused a small amount of blackening or soot marks.
Patient Sequence No: 1, Text Type: N, H10
[21459906]
The abbott m2000rt is part of the m2000 realtime system and is for real-time pcr amplification and detection of nucleic acids. A customer in (b)(6), reported an error code display during initialization of the m2000rt instrument. The abbott field service engineer visited the customer site as a follow up to the customer report and observed a charred mark and some damaged components on the block drive assembly interface. The fse replaced the block drive assembly. There was no injury caused by this incident.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 3005248192-2012-00015 |
MDR Report Key | 2840231 |
Report Source | 07 |
Date Received | 2012-05-01 |
Date of Report | 2012-04-02 |
Date of Event | 2012-03-27 |
Date Mfgr Received | 2012-04-02 |
Date Added to Maude | 2012-12-03 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | JEAN LEETE |
Manufacturer Street | 1300 EAST TOUHY AVE. |
Manufacturer City | DES PLAINES IL 60018 |
Manufacturer Country | US |
Manufacturer Postal | 60018 |
Manufacturer Phone | 2123617274 |
Manufacturer G1 | ABBOTT MOLECULAR, INC. |
Manufacturer Street | 1300 E. TOUHY AVE. |
Manufacturer City | DES PLAINES IL 60018 |
Manufacturer Country | US |
Manufacturer Postal Code | 60018 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ABBOTT M2000RT |
Generic Name | REAL TIME NUCLEIC ACID AMPLIFICATION SYSTEM |
Product Code | OOI |
Date Received | 2012-05-01 |
Catalog Number | 09K15-01 |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MOLECULAR, INC. |
Manufacturer Address | 1300 E. TOUHY AVE. DES PLAINES IL 60018 US 60018 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-05-01 |