CAPTURE-R SELECT 0006446

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2012-11-21 for CAPTURE-R SELECT 0006446 manufactured by Immucor, Inc..

Event Text Entries

[21402255] The customers submitted sample resulted as positive (2+ incompatible) with three submitted donor samples and invalid with one submitted donor sample due to a monolayer failure. Known anti-jkb positive and negative samples were crossmatched with the four submitted donor samples and the expected positive (3+) and negative results were obtained. The crossmatch assay was repeated three more times with the submitted sample; it was not possible to obtain a crossmatch result for sample (b)(4) with donor (b)(6) due to monolayer failures.
Patient Sequence No: 1, Text Type: N, H10


[21415103] A customer reported obtaining an unexpected negative reaction with the crossmatch assay with a known anti-jkb sample.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1034569-2012-00215
MDR Report Key2841353
Report Source01,05,06
Date Received2012-11-21
Date of Report2012-11-21
Date of Event2012-10-04
Date Mfgr Received2012-10-24
Device Manufacturer Date2012-06-10
Date Added to Maude2013-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR SCOTT WEBBER
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 300915625
Manufacturer CountryUS
Manufacturer Postal300915625
Manufacturer Phone7704412051
Manufacturer G1IMMUCOR, INC.
Manufacturer Street3130 GATEWAY DRIVE
Manufacturer CityNORCROSS GA 30091562
Manufacturer CountryUS
Manufacturer Postal Code30091 5625
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCAPTURE-R SELECT
Generic NameBLOOD BANK SUPPLY
Product CodeKSS
Date Received2012-11-21
Catalog Number0006446
Lot NumberSC208
Device Expiration Date2013-06-10
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerIMMUCOR, INC.
Manufacturer Address3130 GATEWAY DRIVE NORCROSS GA 30091562 US 30091 5625


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.