ARCHITECT CYCLOSPORINE 01L75-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01 report with the FDA on 2012-11-21 for ARCHITECT CYCLOSPORINE 01L75-25 manufactured by Abbott Laboratories.

Event Text Entries

[3044219] The customer stated cyclosporine results were out of range low on an architect i1000sr analyzer. There was no adverse impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[10297223] (b)(4). The cause of the archtiect cyclosporine imprecision is due to an interaction between the architect cyclosporine assay and the resin used to produce certain architect reaction vessel (rv) lots. Assay precision can be impacted if rvs containing different resin are mixed in the architect hopper and used to test architect cyclosporine. All architect cyclosporine customers will be instructed to only run architect cyclosporine using rvs manufactured with resin from the same supplier. Rv lots beginning with lot 06083p100 and higher can be used together. Rv lots lower than 06083p100 can be used together.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2012-02080
MDR Report Key2842669
Report Source01
Date Received2012-11-21
Date of Report2012-10-14
Date Mfgr Received2012-11-08
Date Added to Maude2012-11-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 60064350
Manufacturer CountryUS
Manufacturer Postal Code60064 3500
Single Use3
Remedial ActionNO
Previous Use Code3
Removal Correction Number1415939-10/10/11-002-C
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT CYCLOSPORINE
Product CodeMKW
Date Received2012-11-21
Catalog Number01L75-25
Lot Number14603M500
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.