TRANSMOTION MEDICAL, INC. TMM4-XWTB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-11-13 for TRANSMOTION MEDICAL, INC. TMM4-XWTB manufactured by Transmotion Medical Inc..

Event Text Entries

[3044751] Very little information was known about this specific event as multiple contacts were made at the facility around the original occurrence of this incident, and little information was shared at that time. Patient moved up toward the head of the chair, and it started to tip over. The chair was "caught" by two caregivers and moved the patient back down to the seat section. No injury was reported.
Patient Sequence No: 1, Text Type: D, B5


[10317673] Specific serial number unknown by the facility. A general view of all the devices found that nothing at all was found to be deficient with them. Multiple follow-ups were made by our sales representative at the time and little more information was received. Not much else was required as the patient's weight was moved too far up the back section and the warning labels weren't adhered to.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2012-00007
MDR Report Key2845007
Report Source06
Date Received2012-11-13
Date of Report2012-11-12
Date of Event2009-08-07
Date Mfgr Received2009-08-07
Device Manufacturer Date2008-01-01
Date Added to Maude2012-12-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL, INC.
Generic NameTMM5 MOBILE SURGICAL STRETCHER-CHAIR
Product CodeGBB
Date Received2012-11-13
Model NumberTMM4-XWTB
Catalog NumberTMM4-XWTB
Lot Number1822-1833
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2012-11-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.