PET ADVANCE GANTRY 46-311001 NA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2000-06-29 for PET ADVANCE GANTRY 46-311001 NA manufactured by Ge Medical Systems.

Event Text Entries

[167441] Manufacturing discovered a missing lock nut from the tilt mechanism of the pet advance gantry on some systems. This could potentially lead to failure of the tilt mechanism, which in some gantry orientations might result in unwanted tilt of the gantry toward the pt, possibly resulting in injury. There have been no reports of injury associated with this internal finding.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2126677-2000-00007
MDR Report Key284748
Report Source07
Date Received2000-06-29
Date of Report2000-06-29
Date of Event2000-05-11
Date Mfgr Received2000-05-11
Date Added to Maude2000-07-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSTEVEN KAY
Manufacturer StreetPO BOX 414
Manufacturer CityMILWAUKEE WI 53201
Manufacturer CountryUS
Manufacturer Postal53201
Manufacturer Phone2625134064
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NamePET ADVANCE GANTRY
Generic NameNUCLEAR MEDICINE
Product CodeIYY
Date Received2000-06-29
Model Number46-311001
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityN
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key275556
ManufacturerGE MEDICAL SYSTEMS
Manufacturer Address3000 N GRANDVIEW BLVD WAUKESHA WI 53188 US
Baseline Brand NamePET ADVANCE GANTRY
Baseline Generic NameNUCLEAR MEDICINE
Baseline Model No46-311001
Baseline Catalog NoNA
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
10 2000-06-29