STERICHEK TOTAL CHLORINE REAGENT STRIP *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2000-07-07 for STERICHEK TOTAL CHLORINE REAGENT STRIP * manufactured by Environmental Test Systems.

Event Text Entries

[148172] A test strip for measuring low chlorine levels in water used to prepare solutions for kidney dialysis gave falsely low values in the presence of very high levels of chlorine. The tested water came from a holding tank that had been sanitized with chlorine the night before. On june 27, 2000, a voice mail referred to a "problem" with the new reagent strip for low range chlorine. On june 28, co learned that two persons undergoing kidney dialysis became ill because the water used to make the dialysis solution contained chlorine. The dialysis center personnel checked the solution with a test strip designed to measure higher levels of chlorine and showed that there was chlorine present in the solution. There was intervention by dialysis center personnel and the pts have not suffered permanent injury. They are described as fine. The above info is based on telephone conversations. The dialysis center has not yet provided a written report to the mfr.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1833407-2000-00001
MDR Report Key284977
Report Source06
Date Received2000-07-07
Date of Report2000-07-07
Date of Event2000-06-22
Date Mfgr Received2000-06-28
Device Manufacturer Date2000-04-01
Date Added to Maude2000-07-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDAVID MORRIS
Manufacturer Street23575 C.R. 106
Manufacturer CityELKHART IN 46514
Manufacturer CountryUS
Manufacturer Postal46514
Manufacturer Phone2192622060
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERICHEK TOTAL CHLORINE REAGENT STRIP
Generic NameREAGENT STRIP FOR WATER TESTING
Product CodeDJW
Date Received2000-07-07
Model Number*
Catalog Number*
Lot Number003040
ID Number*
Device Expiration Date2002-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key275782
ManufacturerENVIRONMENTAL TEST SYSTEMS
Manufacturer Address23575 C. R. 106 ELKHART IN 46514 US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2000-07-07

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.