ARCHITECT I2000SR ANALYZER 03M74-02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2012-11-30 for ARCHITECT I2000SR ANALYZER 03M74-02 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[2963214] The customer observed a falsely decreased ck-mb result while using the architect i2000sr analyzer. An initial result of <0. 1 ng/ml and a retest result of 1. 9 ng/ml were generated. There was no adverse impact to patient management reported.
Patient Sequence No: 1, Text Type: D, B5


[10261234] (b)(4). An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[10472673] Further investigation of the customer issue included a review of the complaint text, a review of the instrument service, abbott field service evaluation of the instrument, a search for similar complaints, and a review of labeling. A review of the instrument service history did not identify any additional issues. The abbott field service representative replaced the carousel gear drive. No further issues were identified after the gear was replaced. Tracking and trending did not identify an adverse trend. Labeling was reviewed and found to be adequate. Based on the available information no product deficiency of the architect 21000sr analyzer, list number 03m74, was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2012-00492
MDR Report Key2853226
Report Source05
Date Received2012-11-30
Date of Report2012-11-15
Date of Event2012-11-15
Date Mfgr Received2013-01-16
Device Manufacturer Date2010-08-27
Date Added to Maude2013-02-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT I2000SR ANALYZER
Product CodeJLB
Date Received2012-11-30
Catalog Number03M74-02
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-30

© 2026 FDA.report