DIMENSION VISTA 1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2012-11-30 for DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[2958562] A discordant result for acetaminophen (actm) was obtained on a dimension vista 1500 instrument. The result was high and automatically repeated, resulting in a high result. The result was reported out and questioned by the physician(s). The customer reran the sample on another instrument in the laboratory and obtained a low result which was consistent with patient history. A corrected result report was sent to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant high actm result.
Patient Sequence No: 1, Text Type: D, B5


[10260834] The siemens technical support center (tsc) analyzed the database of the instrument. The tsc found excess of protein in the sample. The cause of the discordant actm result is inknown. The instrument is performing within specifications. Further evaluation of the device is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2012-00138
MDR Report Key2853430
Report Source05,06
Date Received2012-11-30
Date of Report2012-11-05
Date of Event2012-11-01
Date Mfgr Received2012-11-05
Date Added to Maude2013-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. MICHAEL METZ
Manufacturer StreetSIEMENS HEALTHCARE DIAGNOSTICS 511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242223
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use0
Remedial ActionOT
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA 1500
Generic NameIMMUNOASSAY ANALYZER
Product CodeLDP
Date Received2012-11-30
Model NumberDIMENSION VISTA 1500
Lot Number12205BE
OperatorHEALTH PROFESSIONAL
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.