MAUDE MDR 2854339

MDR report key
2854339
Report number
2020676-2012-00051
Event key
0
Event type
3
Date of event
2012-10-24
Date received
2012-11-28
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
0

Manufacturer Contact#

Address
4225 EAST LA PALMA AVE. ANAHEIM CA 92807 US
Phone
714-714-7145
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1CHAMBER, HYPERBARICHYPERBARIC MONOPLACE CHAMBERSECHRIST INDUSTRIES, INC.CBF3200R3200RY

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12012-11-280

Event Narratives#

D

Patient 1

DURING SCHEDULED TESTING THE TECH NOTED THAT THE CHAMBER IS COMPRESSING OUTSIDE OF MFR SPECIFICATIONS. SET AT A RATE OF 1, THE CHAMBER IS COMPRESSING AT 1 PSI PER 43 SECONDS. THIS RATE IS FASTER THAN THE SPECIFIED RATE.

N

Patient 1

THE CHAMBER HAD PREVIOUSLY RECEIVED ITS ANNUAL SERVICE MAINTENANCE ON (B)(4) 2012. NO ISSUES REGARDING THE FUNCTIONALITY WERE NOTED ON THE REPORT. EVALUATION OF THE CHAMBER VERIFIED THE REPORTED FAILURE. AT A RATE OF 1 THE CHAMBER WAS COMPRESSING 1PSI PER 44 SECONDS. THE RATE VALVE WAS REPLACED AND THE CHAMBER WAS PERFORMANCE VERIFIED. IT WAS FOUND TO FUNCTION PER FACTORY SPECIFICATION. A REQUEST FOR THE ORIGINAL RATE VALVE HAS BEEN MADE. AN UPDATE WILL BE SENT UPON EVAL OF THE RATE VALVE.