CUFFLATOR ENDOTRACHEAL TUBE INFLATOR AND MANOMETER 8199

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2012-11-02 for CUFFLATOR ENDOTRACHEAL TUBE INFLATOR AND MANOMETER 8199 manufactured by J.t. Powey Company.

Event Text Entries

[2952168] The customer reported the cufflator does not hold pressure. The customer did not specify what date the issue was discovered. There was no pt incident or injury reported.
Patient Sequence No: 1, Text Type: D, B5


[10230417] Results: evaluation of the returned product confirmed the reported issue, needle moves up but does not hold pressure. No readings can be taken since the cufflator does not hold pressure. Rubber protective ring has cuts going around its circumference. Note: the instructions for use states: the cufflator should be calibrated annually, or if measurements fall outside of a range, or if the cufflator needle does not indicate a reading of zero when nothing is connected, or if the unit is ever dropped. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2020362-2012-00638
MDR Report Key2860228
Report Source01,08
Date Received2012-11-02
Date of Report2012-10-15
Date Mfgr Received2012-10-15
Date Added to Maude2013-02-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPAM WAMPLER, ADMINISTRATOR
Manufacturer Street5635 PECK RD.
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal91006
Manufacturer Phone6264433143
Manufacturer G1J. T. POSEY CO.
Manufacturer Street5635 PECK RD.
Manufacturer CityARCADIA CA 91006
Manufacturer CountryUS
Manufacturer Postal Code91006
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCUFFLATOR ENDOTRACHEAL TUBE INFLATOR AND MANOMETER
Product CodeBSK
Date Received2012-11-02
Returned To Mfg2012-10-26
Model Number8199
Catalog Number8199
Lot Number08026457
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerJ.T. POWEY COMPANY
Manufacturer AddressARCADIA CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-02