FIELD SURGICAL LIGHT 2410 MB *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2000-07-17 for FIELD SURGICAL LIGHT 2410 MB * manufactured by Getinge/castle, Inc..

Event Text Entries

[183328] Unit does not stop charging the lead acid batteries when full charge is reached. Batteries get very hot and swell, which ruins the batteries and requires their replacement. Mfr's manual states that batteries require 8 hrs to reach full charge but does not warn that unit must be unplugged. As reported by biomedical.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1019298
MDR Report Key286100
Date Received2000-07-17
Date of Report2000-06-29
Date Added to Maude2000-07-20
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameFIELD SURGICAL LIGHT
Generic NameFIELD SURGICAL LIGHT
Product CodeKYT
Date Received2000-07-17
Model Number2410 MB
Catalog Number*
Lot Number*
ID NumberNSN 6530-01-343-2033
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key276849
ManufacturerGETINGE/CASTLE, INC.
Manufacturer Address7371 SPARTAN BLVD. EAST NORTH CHARLESTON SC 294188444 US


Patients

Patient NumberTreatmentOutcomeDate
10 2000-07-17