WATERS QUATTRO PREMIER M94630DC1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2012-12-05 for WATERS QUATTRO PREMIER M94630DC1 manufactured by Waters Corp..

Event Text Entries

[2951313] The user reported that they found the system shut down when they arrived in the morning. The customer said the rotary interlock box was burned and that there were burn marks on the floor.
Patient Sequence No: 1, Text Type: D, B5


[10331378] Based on information reported by the customer, the instrument was not functioning when they arrived in the morning. There were burn marks on the floor, but the lab sprinklers did not go off. No injuries were reported. The rotary interlock switchbox from the device was returned to waters. A supplemental mdr will be sent when the results of the investigation of the returned component are completed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616673-2012-00003
MDR Report Key2862172
Report Source01,06
Date Received2012-12-05
Date of Report2012-11-09
Date of Event2012-11-09
Date Mfgr Received2012-11-09
Device Manufacturer Date2004-06-01
Date Added to Maude2013-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactMICHAEL PACHECO
Manufacturer Street34 MAPLE ST.
Manufacturer CityMILFORD MA 01757
Manufacturer CountryUS
Manufacturer Postal01757
Manufacturer Phone5084822532
Manufacturer G1WATERS TECHNOLOGIES IRELAND, LTD.
Manufacturer StreetWEXFORD BUSINESS PARK, DRINAGH
Manufacturer CityWEXFORD
Manufacturer CountryEI
Single Use3
Remedial ActionOT
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameWATERS QUATTRO PREMIER
Generic NameDOP MASS SPECTROMETER, CLINICAL USE
Product CodeDOP
Date Received2012-12-05
Returned To Mfg2012-11-30
Model NumberQUATTRO PREMIER
Catalog NumberM94630DC1
Lot NumberNA
ID NumberNA
OperatorOTHER
Device AvailabilityY
Device AgeNA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerWATERS CORP.
Manufacturer Address34 MAPLE ST. MILFORD MA 01757 US 01757


Patients

Patient NumberTreatmentOutcomeDate
10 2012-12-05

© 2026 FDA.report