MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-12-14 for 120 RELIANCE CART WASHER manufactured by Steris Canada Corporation.
[3180079]
The user facility reported that their 120l cart washer was leaking steam. The steam activated the user facility's sprinkler system and water was released. The fire department was dispatched but the user facility was not evacuated as the event occurred overnight. No injuries or procedural delays/cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5
[10485890]
A steris service technician inspected the washer and found that the water booster tank had overheated. This was due to the steam valve having a small cut in the seat. This allowed steam to leak through subsequently setting off the sprinkler system. The technician stated that the unit was left on in an idle state overnight and that the water booster tank is connected to the user facility's steam supply line. The technician repaired the unit, ran a test cycle and confirmed it to be operational.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9680353-2012-00122 |
MDR Report Key | 2873932 |
Report Source | 06 |
Date Received | 2012-12-14 |
Date of Report | 2012-12-14 |
Date of Event | 2012-11-17 |
Date Mfgr Received | 2012-11-17 |
Date Added to Maude | 2013-03-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Reporter Occupation | OTHER HEALTH CARE PROFESSIONAL |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MS LINDSEY MCGOWAN |
Manufacturer Street | 5960 HEISLEY ROAD |
Manufacturer City | MENTOR OH 44060 |
Manufacturer Country | US |
Manufacturer Postal | 44060 |
Manufacturer Phone | 4403927519 |
Manufacturer G1 | STERIS CANADA CORPORATION |
Manufacturer Street | 490, ARMAND-PARIS |
Manufacturer City | QUEBEC, QUEBEC GIC8A3 |
Manufacturer Country | CA |
Manufacturer Postal Code | GIC 8A3 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | 120 RELIANCE CART WASHER |
Generic Name | WASHER |
Product Code | LDS |
Date Received | 2012-12-14 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STERIS CANADA CORPORATION |
Manufacturer Address | 490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2012-12-14 |