WAVESCAN WAVEFRONT SYSTEM WS1 0070-0150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,07 report with the FDA on 2012-12-20 for WAVESCAN WAVEFRONT SYSTEM WS1 0070-0150 manufactured by Abbott Medical Optics.

Event Text Entries

[3101593] The amo field service engineer returned the customer's wavescan computer to amo for evaluation into the computer not functioning. Service depot evaluated the computer and when the power supply was powered up, smoke was observed being emitted from the power supply and the computer was non functional.
Patient Sequence No: 1, Text Type: D, B5


[10388369] The cause of the computer's power supply components failing and emitting smoke was not determined. The computer was repaired with a new power supply installed and was tested. The computer operated correctly following repair. All pertinent information available to amo has been submitted.?
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2012-00191
MDR Report Key2882459
Report Source00,07
Date Received2012-12-20
Date of Report2012-12-18
Date of Event2012-08-14
Date Mfgr Received2012-12-18
Device Manufacturer Date2001-08-02
Date Added to Maude2012-12-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationSERVICE AND TESTING PERSONNEL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAROL KAIL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478598
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer StreetAMO MANUFACTURING 510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN WAVEFRONT SYSTEM WS1
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2012-12-20
Model NumberWS1
Catalog Number0070-0150
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA


Patients

Patient NumberTreatmentOutcomeDate
10 2012-12-20