VA LOCKSCR ?2.4 SELF-TAP L18 TAN 04.210.118

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2012-12-21 for VA LOCKSCR ?2.4 SELF-TAP L18 TAN 04.210.118 manufactured by Synthes Monument.

Event Text Entries

[3098725] A device report from (b)(6) indicated a hospital in (b)(6) reported: screws in unopened packages were taken from stock and it was noticed there were visible chips on all screw heads. There was no patient involvement. This is 6 of 6 reports for this event.
Patient Sequence No: 1, Text Type: D, B5


[10292667] Device used for treatment and not diagnosis. The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.
Patient Sequence No: 1, Text Type: N, H10


[17324314] This device is used for treatment not diagnosis. The screw received is contained in a properly sealed synthes package with no evidence of tears or openings of any type. The printed label indicates that it contains a 04. 210. 118, lot number 6971717. Upon examination through the packaging with a 10x magnifier, chip wrap was found at the neck of the screw. The package was opened to allow further examination. No other burrs/chips were noted. Dhr was reviewed with no complaint related anomalies noted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1719045-2012-01433
MDR Report Key2883320
Report Source01,07
Date Received2012-12-21
Date of Report2012-11-22
Date Mfgr Received2013-01-10
Date Added to Maude2013-03-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSHERRY LAING
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES MONUMENT
Manufacturer Street1101 SYNTHES AVE
Manufacturer CityMONUMENT CO 80132
Manufacturer CountryUS
Manufacturer Postal Code80132
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVA LOCKSCR ?2.4 SELF-TAP L18 TAN
Generic NameVA LOCKSCREW
Product CodeODX
Date Received2012-12-21
Returned To Mfg2012-12-10
Catalog Number04.210.118
Lot Number6971717
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES MONUMENT
Manufacturer Address1101 SYNTHES AVE MONUMENT CO 80132 US 80132


Patients

Patient NumberTreatmentOutcomeDate
10 2012-12-21

© 2026 FDA.report