MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-12-21 for URISYS 1100 03617556690 manufactured by Roche Diagnostics.
[3102861]
Reporter stated that a patient sample was tested on the suspect urisys 1100 which produced a false negative result for protein. An additional test was reportedly performed, using the same strip lot, on a different urisys instrument which reported a protein result of 500 mg/dl. No adverse event was reported. The manufacturer requested the return of the suspect product for evaluation.
Patient Sequence No: 1, Text Type: D, B5
[10367046]
It is not known if the initial reporter has reported or intends to report the event to f. D. A.
Patient Sequence No: 1, Text Type: N, H10
| Report Number | 1823260-2012-06502 |
| MDR Report Key | 2883756 |
| Report Source | 06 |
| Date Received | 2012-12-21 |
| Date of Report | 2013-02-01 |
| Date of Event | 2012-11-28 |
| Date Mfgr Received | 2012-11-28 |
| Date Added to Maude | 2013-03-25 |
| Event Key | 0 |
| Report Source Code | Manufacturer report |
| Manufacturer Link | Y |
| Number of Patients in Event | 0 |
| Adverse Event Flag | 3 |
| Product Problem Flag | 3 |
| Reprocessed and Reused Flag | 3 |
| Health Professional | 3 |
| Initial Report to FDA | 3 |
| Report to FDA | 0 |
| Event Location | 0 |
| Manufacturer Contact | NA JENNIFER WOLFGRAM |
| Manufacturer Street | 9115 HAGUE ROAD NA |
| Manufacturer City | INDIANAPOLIS IN 46250 |
| Manufacturer Country | US |
| Manufacturer Postal | 46250 |
| Manufacturer Phone | 3175217008 |
| Manufacturer G1 | ROCHE DIAGNOSTICS GMBH |
| Manufacturer Street | SANDHOFERSTRASSE 116 NA |
| Manufacturer City | MANNHEIM (BADEN-WURTTEMBERG) 68305 |
| Manufacturer Country | GM |
| Manufacturer Postal Code | 68305 |
| Single Use | 3 |
| Previous Use Code | 3 |
| Removal Correction Number | NA |
| Event Type | 3 |
| Type of Report | 3 |
| Brand Name | URISYS 1100 |
| Generic Name | AUTOMATED URINE ANALYZER |
| Product Code | KHE |
| Date Received | 2012-12-21 |
| Model Number | NA |
| Catalog Number | 03617556690 |
| Lot Number | NA |
| ID Number | NA |
| Device Availability | N |
| Device Age | DA |
| Device Eval'ed by Mfgr | R |
| Device Sequence No | 1 |
| Device Event Key | 0 |
| Manufacturer | ROCHE DIAGNOSTICS |
| Manufacturer Address | 9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250 |
| Patient Number | Treatment | Outcome | Date |
|---|---|---|---|
| 1 | 0 | 2012-12-21 |