OLYMPUS SCOPE CABLE EXERA II MAJ-1430

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2012-11-20 for OLYMPUS SCOPE CABLE EXERA II MAJ-1430 manufactured by Olympus Medical Systems Corporation.

Event Text Entries

[3099215] Olympus was informed that the referenced device failed during a procedure. There was no further detail provided.
Patient Sequence No: 1, Text Type: D, B5


[10363222] Olympus followed-up with the user facility to obtain add'l info regarding this report. The user facility reported that there was a complete loss of image during an unspecified colonoscopy procedure. The intended procedure was said to have been completed with a new cable. There was no pt harm reported. The device eval did not confirm the user's report. The referenced device was evaluated and the image was found flickering when the cable was manipulated at the round connector side. The reference device was evaluated with a (b)(4) cable tester and it passed the continuity check. There was an unstable connection internally with the wires. The referenced device had been forwarded to the original equipment mfr (oem) for further eval.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010047-2012-00434
MDR Report Key2885311
Report Source06
Date Received2012-11-20
Date of Report2012-10-23
Date Mfgr Received2012-10-23
Date Added to Maude2013-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVENUE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Street2951 ISHIKAWA-CHO
Manufacturer CityHACHIOJI-SHI, TOKYO 192-8507
Manufacturer CountryJA
Manufacturer Postal Code192-8507
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS SCOPE CABLE EXERA II
Generic NameSCOPE CABLE
Product CodeKQM
Date Received2012-11-20
Returned To Mfg2012-10-23
Model NumberMAJ-1430
Catalog NumberMAJ-1430
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS MEDICAL SYSTEMS CORPORATION
Manufacturer Address2951 ISHIKAWA-CHO HACHIOJI-SHI, TOKYO 192-8507 JA 192-8507


Patients

Patient NumberTreatmentOutcomeDate
10 2012-11-20

© 2026 FDA.report