OLYMPUS MH-706 MONITOR TRAY 7284292

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2000-08-10 for OLYMPUS MH-706 MONITOR TRAY 7284292 manufactured by Keymed (medical & Industrial Equipment), Ltd..

Event Text Entries

[181929] The head of medical physics called keymed to report that a 20" tv monitor, complete with its support column, had become detached from an olympus wm-30 mobile workstation when the monitor had struck an overhead theatre light gantry while being moved between departments. Examination showed that all four of the m8 x 55mm screws which should secure the support column to the top tray of the wm-30 had worked loose from the unit, leaving it superficially fixed by only two m3 screws, which were not designed to support the weight of the monitor. The monitor fell down the front of the wm-30, attached only by signal cables at the rear, causing the workstation to topple over. No one suffered serious injury in the incident, but one member of staff sustained a cut to the left arm when staff tried to prevent the monitor from falling.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number9611174-2000-00001
MDR Report Key290227
Report Source06
Date Received2000-08-10
Date of Report2000-08-09
Date of Event2000-06-07
Date Mfgr Received2000-06-07
Device Manufacturer Date1999-07-01
Date Added to Maude2000-08-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactMR R GRAY
Manufacturer StreetSTOCK ROAD
Manufacturer CitySOUTHEND-ON-SEA, ESSEX SS2 5QH
Manufacturer CountryUK
Manufacturer PostalSS2 5QH
Manufacturer Phone702616333
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionNO
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS
Generic NameMOBILE ENDOSCOPY WORKSTATION
Product CodeKHX
Date Received2000-08-10
Returned To Mfg2000-06-14
Model NumberMH-706 MONITOR TRAY
Catalog Number7284292
Lot Number113644
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key280884
ManufacturerKEYMED (MEDICAL & INDUSTRIAL EQUIPMENT), LTD.
Manufacturer AddressSTOCK RD. SOUTHEND-ON-SEA, ESSEX UK SS2 5QH
Baseline Brand NameOLYMPUS
Baseline Generic NameMOBILE ENDOSCOPY WORKSTATION
Baseline Model NoMH-706 MONITOR
Baseline Catalog No7284292
Baseline ID*