RELIANCE 120 CART WASHER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-01-10 for RELIANCE 120 CART WASHER manufactured by Steris Canada Corporation.

Event Text Entries

[3305233] The user facility reported that water was leaking from their reliance series 120 washer. The leak was contained by hospital staff and cleaned up using a wet vac. No procedural delays/cancellations have been reported. No injuries to the hospital staff or patients reported.
Patient Sequence No: 1, Text Type: D, B5


[10396240] On (b)(4) 2012, routine preventative maintenance was performed on the device. The steris service technician reported there was an internal leak at the circulation pump. As the service technician's investigation could not be completed at the time of pm, the unit was taken out of service while investigating the root cause of the leak. During this period, unbeknownst to steris, the user returned the unit to service, at which time the reported event occurred. On (b)(6) 2012, the steris technician arrived on site and found water leaking from the unit. The technician inspected the unit and found the leak to be coming from the circulation pump and piping of the unit. The technician replaced the pump hose and seals on the unit. A test cycle was run and the unit was found to be operational.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2013-00002
MDR Report Key2907947
Report Source06
Date Received2013-01-10
Date of Report2013-01-10
Date of Event2012-12-13
Date Mfgr Received2012-12-13
Date Added to Maude2013-04-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS LINDSEY MCGOWAN
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927519
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRELIANCE 120 CART WASHER
Generic NameWASHER
Product CodeLDS
Date Received2013-01-10
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2013-01-10

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.