PATCH 0041

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-01-10 for PATCH 0041 manufactured by Historical Cpi St. Paul.

Event Text Entries

[3305717] Boston scientific received information that this large epicardial tachycardia patch is part of an implanted system that exhibited a high out of range shock impedance measurement one day after implant. This new icd (model/serial f161/(b)(4)) was implanted with another epicardial tachycardia lead (model/serial 0041/(b)(4)) with an adaptor for each lead (model/serial 4312/(b)(4) and 4312/(b)(4)). The cause of the high shock impedance measurement was not unknown. A revision procedure was scheduled and the patient was hospitalized.
Patient Sequence No: 1, Text Type: D, B5


[10351026] Additional information is expected with regard to this issue. This investigation will be updated should additional information become available.
Patient Sequence No: 1, Text Type: N, H10


[18254637] Field follow-up confirmed a fractured lead component. During surgical intervention, the lead was removed from service and another boston scientific right ventricular lead implanted. No resulting adverse patient effects were reported.
Patient Sequence No: 1, Text Type: D, B5


[18409884] No return of product is expected. As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2124215-2012-14580
MDR Report Key2908279
Report Source07
Date Received2013-01-10
Date of Report2012-10-23
Date of Event2012-10-11
Date Mfgr Received2012-10-23
Date Added to Maude2013-01-10
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactSHARON ZURN
Manufacturer Street4100 HAMLINE AVE. N
Manufacturer CityST. PAUL MN 55112
Manufacturer Postal55112
Manufacturer Phone6515824786
Manufacturer G1HISTORICAL CPI ST. PAUL
Manufacturer StreetGUIDANT CORPORATION
Manufacturer CitySAINT PAUL MN
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePATCH
Generic NameIMPLANTABLE LEAD
Product CodeNHW
Date Received2013-01-10
Model Number0041
ID NumberL 67
Device Expiration Date1992-09-25
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerHISTORICAL CPI ST. PAUL
Manufacturer AddressGUIDANT CORPORATION SAINT PAUL MN


Patients

Patient NumberTreatmentOutcomeDate
143121. Hospitalization; 2. Required No Informationntervention 2013-01-10