ENDURANCE PYRAMID CANE W1300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2006-09-29 for ENDURANCE PYRAMID CANE W1300 manufactured by R. Poon Medical Products Co., Ltd..

Event Text Entries

[21985836] According to the user's wife, the incident occurred at around 8:30am on friday (b)(6) 2006. The user was entering the bathroom and she heard a noise that sounded like a fall. When she entered the bathroom, she found the user had fallen to the floor and the broken cane was next to him. The user claims that he fell when the outside left leg broke. The user was very dazed by the incident and had to lie on the floor for approximately 15 minutes. The user was taken to the doctor by his wife. The user's wife claimed that he was bruised and cut as a result of the incident, but provided no physical evidence to support the claim.
Patient Sequence No: 1, Text Type: D, B5


[22201793] The reporter sent pictures of the device to the distributor for eval. It appears that the outside left leg broke where the crossbar is attached. We (the distributor of the device) requested that user return the device for eval; however, the user did not want to return it. We are unable to further determine the cause of the malfunction at this time because we do not have possession of the cane involved in the incident. No similar incidents were reported to the mfr. We will return the device to the mfr for testing if it is returned by the user. In reviewing photo taken by the dealer and sent to us, both outside legs of the cane are bent inward, tending to indicate that at least one leg was wedged or caught before the fall. No corroborating evidence is available as use was alone in the bathroom when the fall occurred.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1000282279-2006-00011
MDR Report Key2923268
Report Source08
Date Received2006-09-29
Date of Report2006-09-05
Date of Event2006-09-01
Date Mfgr Received2006-09-07
Device Manufacturer Date2004-06-01
Date Added to Maude2013-01-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetNEW CITY DISTRICT, DANZAO NANHAI, FOSHAN
Manufacturer CityGUANGDONG 528216
Manufacturer CountryCH
Manufacturer Postal528216
Manufacturer Phone5785448866
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameENDURANCE PYRAMID CANE
Generic NameKHY CANE, SAFETY WALK
Product CodeKHY
Date Received2006-09-29
Model NumberW1300
Catalog NumberW1300
Lot NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerR. POON MEDICAL PRODUCTS CO., LTD.
Manufacturer AddressGUANGDONG CH


Patients

Patient NumberTreatmentOutcomeDate
10 2006-09-29