MALYUGIN RING SYSTEM MAL-0002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-01-16 for MALYUGIN RING SYSTEM MAL-0002 manufactured by Microsurgical Technology, Inc..

Event Text Entries

[3371798] When the surgeon attempted to remove a malyugin ring a small strand of iris became tangled in the ring. The surgeon thought the small strand would tear easily and tugged on the iris causing a tear and bleeding. A second attempt to remove the ring with the inserter was successful.
Patient Sequence No: 1, Text Type: D, B5


[10408689] The pt? S iris was bound down with synechia. The surgeon cleared the synechia 360 degrees with viscoat and afterward the iris had significant prolapse with one of the eyelets of the ring protruding into the incision. Upon completing cataract removal the ring was disengaged but the iris had a moth-eaten consistency and a portion of the iris had become entangled in the ring resulting in the event. The surgeon reported the incision was too short and four sutures were used to close the wound. The surgeon expected the blood to clear in a couple of months and the pt to fully recover.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3019924-2013-00002
MDR Report Key2928885
Report Source05
Date Received2013-01-16
Date of Report2013-01-14
Date of Event2012-12-21
Date Added to Maude2013-01-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street8415 154TH AVE., N.E.
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMALYUGIN RING SYSTEM
Generic NameIRIS RETRACTOR
Product CodeHNI
Date Received2013-01-16
Model NumberMAL-0002
Catalog NumberMAL-0002
Lot NumberUNKNOWN
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY, INC.
Manufacturer AddressREDMOND WA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-01-16

© 2024 FDA.report