CLINITEK ADVANTUS

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-01-22 for CLINITEK ADVANTUS manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[3154438] Customer reports false negative leukocyte results on a sample which, upon microscopic exam, is 2+ positive for leukocytes. No report of injury with the event.
Patient Sequence No: 1, Text Type: D, B5


[10447889] Customer reports they had never cleaned the system. Customer was provided instruction on proper cleaning and maintenance of the system. Customer was sent new lot of test strips and will perform correlation study. Mfr awaiting results.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2012-00080
MDR Report Key2931384
Report Source07
Date Received2013-01-22
Date of Report2012-12-03
Date of Event2012-12-03
Date Mfgr Received2012-12-03
Date Added to Maude2013-02-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARK SMITH
Manufacturer Street2 EDGEWATER DRIVE
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693158
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetNORTHERN ROAD
Manufacturer CitySUDBURY, SUFFOLK CO102XQ
Manufacturer CountryUK
Manufacturer Postal CodeCO10 2XQ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLINITEK ADVANTUS
Generic NameCLINITEK ADVANTUS
Product CodeCDM
Date Received2013-01-22
Model NumberADVANTUS
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2013-01-22

© 2026 FDA.report