SUPERSTAND HPS-2

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2013-02-04 for SUPERSTAND HPS-2 manufactured by The Standing Company.

Event Text Entries

[3286694] The customer called in on (b)(6) 2012 and said he fell out of his wheelchair around (b)(6). Customer had been using the stand up wheelchair since (b)(6) 2012 (3 months). Customer said he leaned over to pick up keys off the floor and fell out of his chair when the wheelchair rose up as he shifted his weight off the seat because he didn't bring it down to the locked position. Customer got back in the chair and continued with his day. He did not think it was a big deal. Customer did not seek medical care nor see a doctor until two weeks later. The doctor identified a fracture of the leg and advised customer to stay off his leg for a month or so until it healed. The customer notified the standing company on (b)(6) 2013 that his leg has not healed properly and that further medical attention may be necessary. We are currently awaiting for any medical updates.
Patient Sequence No: 1, Text Type: D, B5


[10469735] Manufacturer has reviewed safe operating procedures (sops) with customer again. Customer acknowledges his wheelchair was not down totally in the fail-safe position. Manufacturer updated the safe operating procedures on 12/13/2012. (s. O. P. ) rule #1, now reads: before transferring in or out of the chair or before bending over to reach the floor... I must make sure the safety lock is on (that is, chair is in full down position - with raise levers all the way to the rear). (s. O. P) rule #17, i understand superstand must be completely down before transferring out. (listen for click and clunk when raise levers go all the way to the rear).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007494904-2013-00001
MDR Report Key2949526
Report Source04
Date Received2013-02-04
Date of Report2013-02-04
Date of Event2012-11-22
Date Mfgr Received2012-12-07
Device Manufacturer Date2012-08-26
Date Added to Maude2013-02-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5848 DIXIE HIGHWAY
Manufacturer CitySAGINAW MI 48601
Manufacturer CountryUS
Manufacturer Postal48601
Manufacturer Phone9897469100
Single Use3
Remedial ActionRB
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPERSTAND
Generic NameSTAND UP WHEELCHAIR
Product CodeIPL
Date Received2013-02-04
Model NumberHPS-2
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTHE STANDING COMPANY
Manufacturer AddressSAGINAW MI US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-02-04