IPV-1C *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2013-01-16 for IPV-1C * manufactured by Percussionaire Corporation.

Event Text Entries

[19733128] Respiratory treatment was briefly started (<10 seconds) when high pressures occurred using the intrapulmonary percussive ventilator (ipv) machine. The patient's heart rate dropped to approximately 30 beats-per-minute and immediately returned to normal within 5 seconds. The nurse witnessed the event. Staff further tested the unit and high pressures were noted and the machine was not used further. Called for other respiratory therapists (rt's) to look at the unit and they also failed to see anything obviously wrong with the setup. This ipv medical treatment was skipped as another unit was not available. Clinical engineering states that there was a hard failure inside the machine that caused the high pressure. Machine returned to manufacture. Manufacture repair notes indicate "regulator creeps and regulator ring is broken, needs replaced. Metering valve goes up to fast and slow when pushed on the stem, needs replaced. Adjustable bulk head needs replaced. "======================manufacturer response for ipv machine, ipv-1c (per site reporter). ======================see repair note in event section. What was the original intended procedure? Ipv respiratory treatment.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2950185
MDR Report Key2950185
Date Received2013-01-16
Date of Report2013-01-16
Date of Event2012-11-29
Report Date2013-01-16
Date Reported to FDA2013-01-16
Date Reported to Mfgr2013-02-08
Date Added to Maude2013-02-08
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameIPV-1C
Generic NameDEVICE, POSITIVE PRESSURE BREATHING
Product CodeNHJ
Date Received2013-01-16
Returned To Mfg2013-01-02
Model NumberIPV-1C
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityR
Device Age8 YR
Device Sequence No1
Device Event Key0
ManufacturerPERCUSSIONAIRE CORPORATION
Manufacturer AddressP.O. BOX 817 SANDPOINT ID 83864 US 83864


Patients

Patient NumberTreatmentOutcomeDate