ARCHITECT CYCLOSPORINE 01L75-25

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-02-11 for ARCHITECT CYCLOSPORINE 01L75-25 manufactured by Abbott Laboratories.

Event Text Entries

[3137218] The customer observed discrepant cyclosporine results while using the architect cyclosporine assay. The customer provided the following results (ng/ml) sid (b)(6): initial 332. 6, retest 108; compared to mass spec results: 62, 71 no impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


[10459256] (b)(4). An evaluation is in process. A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[10727892] Further investigation of the customer issue included a review of the complaint text, in-house testing, a search for similar complaints, and a review of labeling. It was noted that sample handling issues were identified at the customer site. An accuracy testing protocol was executed using lot 18265m500; testing met the acceptance criteria and determined the reagent is performing acceptably. Tracking and trending did not identify an adverse trend. Labeling was reviewed and found to be adequate. Based on the available information no product deficiency and no malfunction of the architect cyclosporine reagent list number 01l75, lot number 18265m500, was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2013-00037
MDR Report Key2955292
Report Source05
Date Received2013-02-11
Date of Report2013-01-24
Date of Event2013-01-23
Date Mfgr Received2013-04-19
Device Manufacturer Date2012-09-01
Date Added to Maude2013-02-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 60064350
Manufacturer CountryUS
Manufacturer Postal Code60064 3500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT CYCLOSPORINE
Product CodeMKW
Date Received2013-02-11
Catalog Number01L75-25
Lot Number18265M500
Device Expiration Date2013-07-30
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-11

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.