HIGHT THROUGHTPUT AUTOMATION

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-02-06 for HIGHT THROUGHTPUT AUTOMATION manufactured by Hologic.

Event Text Entries

[15016255] Upon starting instrument customer experienced a strong burning rubber smell, called technical support, and turned off the instrument. No person was injured in this event and there was no delay in or impact to patient diagnosis. Customer confirmed that no medical attention was needed by laboratory technicians with respect to this event. According to the supplier of the component involved in this event, the circuit board involved was developed in compliance with (b)(4) and the requirement that overheated pcb may not burn and cause spreading fire was fulfilled and confirmed by in-house testing. Root cause of this event is under investigation.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2013-00024
MDR Report Key2958086
Report Source05
Date Received2013-02-06
Date of Report2013-01-10
Date of Event2013-01-10
Date Added to Maude2013-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRANDALL COVILL, MGR.
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5082638853
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHIGHT THROUGHTPUT AUTOMATION
Generic NameHTA
Product CodeMAQ
Date Received2013-02-06
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC
Manufacturer AddressMARLBOROUGH MA US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-06

© 2026 FDA.report