OLYMPIC PASTEURMATIC 3000 007169 52640

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-02-08 for OLYMPIC PASTEURMATIC 3000 007169 52640 manufactured by Natus Medical Inc.

Event Text Entries

[3135779] The heaters inside the device stayed on after the customer turned off the power. Later, when the customer returned to the unit, the heaters were engaged (there was no water in the system). The device showed a configuration error stating "heating indicated". The front basket shield melted and the heaters and heater breakers turned purple. The main breaker switch was turned off to disengage the heaters. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10459375] The cause of the malfunction is most likely a faulty k2 relay. Natus technical service requested that the biomed department turn the main breaker back on and fill the device halfway with water until the k1 relay engaged. They powered off the device and the k1 relay stayed engaged. Once biomed replaced the k2 relay, the k1 relay operated correctly. Suspect relay has been returned to natus for eval, along with a printout of the last 50 cycles from the device. Experiments will be run to try to simulate the event. A follow up report will be submitted when add'l info is available.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2013-00002
MDR Report Key2961303
Report Source05
Date Received2013-02-08
Date of Report2013-01-10
Date of Event2013-01-09
Date Mfgr Received2013-01-10
Device Manufacturer Date2012-05-20
Date Added to Maude2013-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTHERESE LEGGIERE
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685173
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC PASTEURMATIC 3000
Generic NamePASTEURIZER
Product CodeLDS
Date Received2013-02-08
Returned To Mfg2013-01-24
Model Number007169
Catalog Number52640
Lot NumberNOT APPLICABLE
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-08

© 2026 FDA.report