DADE ACTIN FSL ACTIVATED PTT REAGENT B4219

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-02-18 for DADE ACTIN FSL ACTIVATED PTT REAGENT B4219 manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[3135982] A discrepant elevated activated partial thromboplastin time (aptt) clotting time result was reported on a patient sample. The same sample was repeated and a lower result was obtained. The patient was treated on the basis of the elevated aptt result. The patient was administered fresh frozen plasma and cryoprecipitates. There is no report of adverse outcome to the patient as a result of the discrepant elevated aptt result and treatment.
Patient Sequence No: 1, Text Type: D, B5


[10572716] The cause of discrepant elevated activated partial thromboplastin time (aptt) results is unknown. A contributor to at least one of the elevated results was user error. One sample was draw from a line with heparin in the line. The pattern of discrepant results is strongly suggestive of patient sample specific factors. The instrument and software is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610806-2013-00004
MDR Report Key2966262
Report Source05,06
Date Received2013-02-18
Date of Report2013-01-25
Date of Event2013-01-24
Date Mfgr Received2013-01-25
Device Manufacturer Date2012-04-05
Date Added to Maude2013-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Generic NameDADE ACTIN FSL ACTIVATED PTT REAGENT
Product CodeGGW
Date Received2013-02-18
Catalog NumberB4219
Lot Number547339
Device Expiration Date2014-03-19
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Address76 EMIL VON BEHRING STRASSE MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-18

© 2026 FDA.report