COBAS E-ELECSYS ESTRADIOL GEN.2 03000079190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-02-19 for COBAS E-ELECSYS ESTRADIOL GEN.2 03000079190 manufactured by Roche Diagnostics.

Event Text Entries

[3191014] The customer received questionable estradiol results for three patient samples and a questionable progesterone result for one other patient sample from the cobas e601 analyzer serial number (b)(4). Of the data provided, the estradiol result for one patient sample was discrepant. The initial result was 866. 5 pg/ml and the repeat result was 599. 5 pg/ml all of the results were reported outside the laboratory. The repeat results were believed to be correct. The patients were not adversely affected. The customer stated they replaced the estradiol reagent pack which resolved the issue. The field service representative determined the reagent was old, expired or exposed to ambient temperature. He confirmed the reagent was replaced with a new lot and problem was no longer encountered.
Patient Sequence No: 1, Text Type: D, B5


[10541389] It was unknown if the initial reporter sent report to the fda.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2013-00990
MDR Report Key2967936
Report Source05,06
Date Received2013-02-19
Date of Report2013-02-19
Date of Event2013-01-29
Date Mfgr Received2013-01-30
Date Added to Maude2013-02-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA JENNIFER WOLFGRAM
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175217008
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameCOBAS E-ELECSYS ESTRADIOL GEN.2
Generic NameRADIOIMMUNOASSAY,ESTRADIOL
Product CodeJIS
Date Received2013-02-19
Model NumberNA
Catalog Number03000079190
Lot Number17033801
ID NumberNA
Device Expiration Date2013-09-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-19

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.