MTS 24 PLACE CENTRIFUGE MTS515060

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2006-10-16 for MTS 24 PLACE CENTRIFUGE MTS515060 manufactured by Micro Typing Systems, Inc..

Event Text Entries

[16909672] The customer reported the mts centrifuge was not operating at the correct speed. Gel cards centrifuged at improper speed could results in erroneous results. No erroneous test results were reported there was no report of harm as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[16951451] Investigation into this event found that the user observed the issue and aborted testing, thereby preventing erroneous results from being reported. The centrifuge was returned to the service depot confirmed the customer's complaint. The root cause of the event is unk. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1056600-2006-00211
MDR Report Key2971445
Report Source05,06
Date Received2006-10-16
Date of Report2006-10-13
Date of Event2005-04-13
Date Mfgr Received2005-04-13
Device Manufacturer Date2001-10-01
Date Added to Maude2013-02-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDAVID HAGEN, SITE MGR
Manufacturer Street1295 S.W. 29TH AVE.
Manufacturer CityPOMPANO BEACH FL 33069
Manufacturer CountryUS
Manufacturer Postal33069
Manufacturer Phone9546239526
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMTS 24 PLACE CENTRIFUGE
Generic NameCENTRIFUGE
Product CodeKSO
Date Received2006-10-16
Model NumberNA
Catalog NumberMTS515060
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICRO TYPING SYSTEMS, INC.
Manufacturer Address1295 S.W. 29TH AVE. POMPANO BEACH FL 33069 US 33069


Patients

Patient NumberTreatmentOutcomeDate
10 2006-10-16

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.