OLYMPIC COLL-CAP SYSTEM 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-02-07 for OLYMPIC COLL-CAP SYSTEM 60010 manufactured by Natus Medical Incorporated.

Event Text Entries

[3335233] An infant was being treated for hypoxic ischemic encephalopathy with the cool-cap system when a pop sound came from the unit as well as a smoke smell during a rewarming process of a child. The unit shut down and the smoke smell was traced to filter below the water bag (power supply). There was no reported pt harm. There was no damage to the outlet or the power cable. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10510083] The cause of this malfunction is most likely a failed power supply. The power supply on this sys is due to be replaced with a new power supply as part of a field corrective action ((b)(4)), but this failure occurred before the svc could be performed. The customer has been shipped a loaner device, and the affected unit has been returned to natus medical's (b)(4) facility for evaluation; and it is confirmed that this unit contains an older, excelsys power supply that is subject of the field corrective action. Natus anticipates no further reporting on this incident unless further investigation ito the failed sys shows a cause other than the excelsys power supply for the reported failure.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2013-00001
MDR Report Key2975399
Report Source05
Date Received2013-02-07
Date of Report2013-01-09
Date of Event2013-01-08
Date Mfgr Received2013-01-09
Device Manufacturer Date2008-03-01
Date Added to Maude2013-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEVE HESLER
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685156
Single Use3
Previous Use Code3
Removal Correction Number3018859-05/25/12-002-C
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COLL-CAP SYSTEM
Generic NameCOOLING CAP (INFANTS)
Product CodeMXM
Date Received2013-02-07
Returned To Mfg2013-02-04
Model Number60010
Catalog Number60010
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-07

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.