ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM 8065750290

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2013-02-18 for ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM 8065750290 manufactured by Alcon - Irvine Technology Center.

Event Text Entries

[21763807] A customer reported that the system turned off on its own during a vitrectomy procedure. After a more than 15 minute delay, the procedure was completed with no harm to the pt.
Patient Sequence No: 1, Text Type: D, B5


[21892057] A review of the service report indicates the system shuts down on its own. The company rep examined the system. The company rep cleaned the air filter and the air pathways. The system was then tested and met all product specifications. There was no sample returned for eval and no additional information provided related to this event. For this reason, steps could not be taken to replicate of confirm the reported event. A review of complaints for the last 24 months did indicate one similar report for this system. The root cause cannot be determined conclusively. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2013-00227
MDR Report Key2976619
Report Source01,05
Date Received2013-02-18
Date of Report2013-01-25
Date of Event2013-01-23
Date Mfgr Received2013-01-25
Device Manufacturer Date2006-08-01
Date Added to Maude2013-03-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTERRY DAGNON
Manufacturer Street6201 SOUTH FREEWAY R7-18
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175514325
Manufacturer G1ALCON - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PKWY.
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM
Generic NameENDOILLUMINATOR
Product CodeMPA
Date Received2013-02-18
Model NumberNA
Catalog Number8065750290
Lot NumberNA
ID Number1.18 - 2.18
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerALCON - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PKWY. IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-18