PRECISION MEDICAL INC. FLOW METER 2MFA1001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-01-23 for PRECISION MEDICAL INC. FLOW METER 2MFA1001 manufactured by Precision Medical Inc..

Event Text Entries

[19687331] Flow meter exploded breaking flow tube. Flow meter was in a unoccupied room, in the off position when the flow tube broke for no apparent reason. No property damage, and no one was injured.
Patient Sequence No: 1, Text Type: D, B5


[19838906] This device is being sent to precision medical for evaluation. Under returned authorization number (b)(4) not received as of (b)(4)2013. The flow meter was returned to precision medical inc on (b)(4) 2013. As found description: the outside housing is broken, with pieces missing, the flow tube itself is missing. The body appears to be whole. The housing has turned yellow from age. The flow meter is 16 years old, with a life expectancy of 10 years. Probable cause of housing failure, was a weakness caused by stress of damage or age. Conclusion: would recommend replacing all flow meters older than ten years of age, or at least an inspection of all flow meters, for damage or aging.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2523148-2013-00001
MDR Report Key2977786
Report Source07
Date Received2013-01-23
Date of Report2013-01-11
Date of Event2013-01-10
Date Facility Aware2013-01-10
Report Date2013-01-11
Date Reported to FDA2013-01-11
Date Reported to Mfgr2013-01-11
Date Mfgr Received2013-01-10
Device Manufacturer Date1997-03-01
Date Added to Maude2013-04-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactJIM PARKER
Manufacturer Street300 HELD DR.
Manufacturer CityNORTHAMPTON PA 18067
Manufacturer CountryUS
Manufacturer Postal18067
Manufacturer Phone6102626090
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRECISION MEDICAL INC. FLOW METER
Generic NameOXYGEN FLOW METER
Product CodeCAX
Date Received2013-01-23
Returned To Mfg2013-01-10
Model Number2MFA1001
Catalog Number2MFA1001
Lot Number0397
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age16 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerPRECISION MEDICAL INC.
Manufacturer Address300 HELD DR. NORTHAMPTON PA 18067 US 18067


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-01-23

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.