OLMPUS VIDEO TELESCOPES A50021A

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-02-01 for OLMPUS VIDEO TELESCOPES A50021A manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[3170337] Olympus was informed that the image went completely dark during an unidentified procedure. There was no report of any pt harm.
Patient Sequence No: 1, Text Type: D, B5


[10568462] Olympus followed-up with the user facility to obtain additional information regarding this report. The user facility reported that the reported phenomenon occurred during a laparoscopic procedure where the users were unable to restore the image after it went completely black. The intended procedure was said to have been completed. The device referenced in this report was returned to olympus for evaluation. The evaluation found the image with spotted dark stain throughout with discoloration, and the image was intermittent. The video connector was severely cracked and the surface of the outer tube was noted with deep scrape and indentation throughout. This report is being submitted as a medical device report in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610773-2013-00021
MDR Report Key2980230
Report Source06
Date Received2013-02-01
Date of Report2013-01-04
Date of Event2012-12-19
Date Mfgr Received2013-01-04
Date Added to Maude2013-04-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLAURA STORMS-TYLER
Manufacturer Street2400 RINGWOOD AVE
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4848965688
Manufacturer G1OLYMPUS WINTER & IBE GMBH
Manufacturer StreetKUEHNSTRASSE 61
Manufacturer CityHAMBURG 22045
Manufacturer CountryGM
Manufacturer Postal Code22045
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLMPUS VIDEO TELESCOPES
Generic NameTELESCOPE
Product CodeFBP
Date Received2013-02-01
Returned To Mfg2013-01-08
Model NumberA50021A
Catalog NumberA50021A
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-01

© 2026 FDA.report