ORTHOVISION TABLE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-02-28 for ORTHOVISION TABLE manufactured by Steris Corporation - Montgomery.

Event Text Entries

[17131559] The user facility reported that when a patient is placed on the orthovision table it would slowly drift down. No injuries or procedural delays/cancellations reported.
Patient Sequence No: 1, Text Type: D, B5


[17358239] A steris service technician inspected the table and found that the table top would drift slowly down due to leakage within the lift cylinder when patient weight was applied. The user facility stated to the steris technician that they were aware of the need to repair the table but continued to use it. The technician stated that they were using some unspecified means of blocking/stopping the table drift. The technician replaced the lift cylinder, tested the table and confirmed it to be operational. The table is not under steris service contract and is serviced and maintained by the user facility.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1043572-2013-00017
MDR Report Key2983211
Report Source06
Date Received2013-02-28
Date of Report2013-02-28
Date of Event2013-01-29
Date Mfgr Received2013-01-29
Date Added to Maude2013-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS LINDSEY MCGOWAN
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927519
Manufacturer G1STERIS CORPORATION - MONTGOMERY
Manufacturer Street2720 GUNTER PARK DRIVE EAST
Manufacturer CityMONTGOMERY AL 36109
Manufacturer CountryUS
Manufacturer Postal Code36109
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameORTHOVISION TABLE
Generic NameTABLE
Product CodeJEB
Date Received2013-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CORPORATION - MONTGOMERY
Manufacturer Address2720 GUNTER PARK DRIVE EAST MONTGOMERY AL 36109 US 36109


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-28

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.