ANODYNE THERAPY 480 498005

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-02-05 for ANODYNE THERAPY 480 498005 manufactured by Anodyne Therapy, Llc.

Event Text Entries

[17348233] Customer reported that the bar graph display did not go to the maximum setting. We requested that the unit be returned for evaluation, and a transistor was found to not be working as intended. There was no adverse event reported.
Patient Sequence No: 1, Text Type: D, B5


[17476466] The system involved in this complaint was returned for evaluation and a transistor failure was detected. No adverse event was reported, however, this type of malfunction could potentially cause an adverse event. The number of incidents of this type remain within the expected rate for this product based upon the number of incidents reported and the number of units in distribution. Company continues to monitor and trend events of this type.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1055581-2013-00001
MDR Report Key2985387
Report Source05
Date Received2013-02-05
Date of Report2013-02-05
Date of Event2012-12-21
Date Mfgr Received2013-01-11
Device Manufacturer Date2005-03-01
Date Added to Maude2013-04-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCANDACE TURTZO
Manufacturer Street14105 MCCORMICK DR.
Manufacturer CityTAMPA FL 33626
Manufacturer CountryUS
Manufacturer Postal33626
Manufacturer Phone8133424432
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameANODYNE THERAPY
Generic NameINFRARED LAMP
Product CodeILY
Date Received2013-02-05
Returned To Mfg2013-01-04
Model Number480
Catalog Number498005
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerANODYNE THERAPY, LLC
Manufacturer AddressTAMPA FL 33626 US 33626


Patients

Patient NumberTreatmentOutcomeDate
10 2013-02-05

© 2026 FDA.report