[15099207] The customer stated that one patient sample (specimen? ) generated higher than expected results of 120 and 140 ng/ml for the architect cyclosporine assay. The patient sample had previously (2 days earlier) generated an expected result of 71 ng/ml. No impact to patient management was reported.
Patient Sequence No: 1, Text Type: D, B5


[15417105] (b)(4). Product evaluation is in process and the results will be submitted in a follow-up report.
Patient Sequence No: 1, Text Type: N, H10


[22339097] Product evaluation was completed in order to investigate this issue. Review of ticket trending and lot search data did not identify atypical complaint activity related to the complaint issue. Accuracy testing was completed using retained kits of reagent lot 16054m500. Acceptance criteria were met, which indicates acceptable product performance. A deficiency was not identified as panel testing shows that the likely cause lot performs per specification. Based on the evaluation results, no malfunction of the lot in question was identified.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1415939-2013-00093
MDR Report Key2991415
Report Source01
Date Received2013-03-06
Date of Report2013-02-14
Date of Event2013-02-14
Date Mfgr Received2013-05-07
Device Manufacturer Date2012-07-01
Date Added to Maude2013-03-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT LABORATORIES
Manufacturer Street100 ABBOTT PARK ROAD
Manufacturer CityABBOTT PARK IL 60064350
Manufacturer CountryUS
Manufacturer Postal Code60064 3500
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameARCHITECT CYCLOSPORINE
Product CodeMKW
Date Received2013-03-06
Catalog Number01L75-25
Lot Number16054M500
Device Expiration Date2013-05-09
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT LABORATORIES
Manufacturer Address100 ABBOTT PARK ROAD ABBOTT PARK IL 60064350 US 60064 3500


Patients

Patient NumberTreatmentOutcomeDate
10 2013-03-06