OLYMPUS INSTRUMENT CHANNEL ADAPTOR MAJ-1606 K10007072&K10016091

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-02-28 for OLYMPUS INSTRUMENT CHANNEL ADAPTOR MAJ-1606 K10007072&K10016091 manufactured by Keymed (medical & Industrial Equipment), Ltd..

Event Text Entries

[3170623] During a diagnostic colonoscopy, when the user inserted biopsy forceps (fb-210u) into the instrument channel adaptor (maj-1606) of an endoscope, the back seal part of the maj-1606 detached and dropped into the pt's body. The part was retrieved using unspecified forceps. There was no pt injury reported.
Patient Sequence No: 1, Text Type: D, B5


[10529999] The maj-1606 is a single use device which is supplied sterile. The device referenced in this report has been returned to olympus (b)(4) who have arranged sterilization of the device. Once sterilization is complete, the device will be returned to (b)(4) to complete the full investigation. (b)(4) have received photographs of the device referenced in this report and are working to replicate the fault on a maj-1606. To assist the investigation, (b)(4) have requested olympus (b)(4) to ask the health center questions relating to the identity of the lot number, the manipulation of the forceps, and the condition of the forceps after the event. There is no reported pt injury and this report is submitted in an abundance of caution.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9611174-2013-00001
MDR Report Key2992921
Report Source01,07
Date Received2013-02-28
Date of Report2013-02-28
Date of Event2013-02-07
Date Mfgr Received2013-02-07
Date Added to Maude2013-05-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactALISON PRIOR, MANAGER
Manufacturer StreetSTOCK RD.
Manufacturer CitySOUTHEND-ON-SEA, ESSEX SS2 5GH
Manufacturer CountryUK
Manufacturer PostalSS2 5GH
Manufacturer Phone1702616333
Manufacturer G1MEDICAL DEVICE MANUFACTURING CENTER
Manufacturer StreetTEMPLE FARM INDUSTRIAL ESTATE
Manufacturer CitySOUTHEND-ON-SEA, ESSEX SS25TF
Manufacturer CountryUK
Manufacturer Postal CodeSS2 5TF
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS INSTRUMENT CHANNEL ADAPTOR
Generic NameMAJ-1606 INSTRUMENT CHANNEL ADAPTOR
Product CodeFEQ
Date Received2013-02-28
Model NumberMAJ-1606
Catalog NumberK10007072&K10016091
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL & INDUSTRIAL EQUIPMENT), LTD.
Manufacturer AddressUK


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-02-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.