ETR COUCH *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2000-10-02 for ETR COUCH * manufactured by Varian Medical Systems, Uk Ltd..

Event Text Entries

[170252] A radiation therapy threatment had been completed when the pt attempted to raise themselves and get off the treatment couch before the therapists had finished lowering it. During this movement the pt gripped the posterior support panel such that as pt sat up they lifted the panel up and moved it off of the channel or groove that holds it in place on the center beam of the couch assembly. Subsequently the panel began to slide to one side, causing the pt to lose their balance and fall to the floor, fracturing their hip.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number2914292-2000-00003
MDR Report Key299350
Report Source05,06
Date Received2000-10-02
Date of Report2000-09-05
Date of Event2000-09-05
Date Mfgr Received2000-09-05
Date Added to Maude2000-10-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactDALE REYNOLDS
Manufacturer Street3100 HANSEN WAY, M/S E-055
Manufacturer CityPALO ALTO CA 94306
Manufacturer CountryUS
Manufacturer Postal94306
Manufacturer Phone6504246773
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameETR COUCH
Generic NamePATIENT SUPPORT ASSEMBLY
Product CodeJAI
Date Received2000-10-02
Model Number*
Catalog Number*
Lot Number*
ID NumberP/N 879225
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key289768
ManufacturerVARIAN MEDICAL SYSTEMS, UK LTD.
Manufacturer AddressGATWICK ROAD CRAWLEY, WEST SUSSEX UK RH10 2RG
Baseline Brand NameETR COUCH
Baseline Generic NamePATIENT SUPPORT ASSEMBLY
Baseline Model No*
Baseline Catalog No*
Baseline IDP/N 879225


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2000-10-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.