ALTERG ANTI-GRAVITY TREADMILL M320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2013-03-05 for ALTERG ANTI-GRAVITY TREADMILL M320 manufactured by Alterg Inc.

Event Text Entries

[3175010] Customer reported that 5 student athletes contracted a staph infection transmitted through the sharing of shorts while using the alterg anti-gravity treadmill. Per the customer: about 5 cases. Diagnosed by a medical doctor. Happens with college age female athletes who have some chafing issues between their legs. The staph makes the problem a lot worse.
Patient Sequence No: 1, Text Type: D, B5


[10518197] Per statements from customer, multiple uses of the shorts garment were happening between wash cycles. The impacted facility hired someone to come in and test the ph of the water. They are confident in the disinfecting at the water temp they are now using (no longer cold water) will resolve the issue. The site has also bought more shorts garments to eliminate the need for sharing of the garments between wash cycles. Alterg will evaluate the labeling instructions for appropriate handling and use of shorts to users; updating instructions as appropriate. Alterg will continue to track this issue in our complaint system for re-occurrence as well as effectiveness of any changes implemented.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007033433-2013-00001
MDR Report Key2994433
Report Source04
Date Received2013-03-05
Date of Report2013-03-05
Date of Event2013-02-14
Date Mfgr Received2013-02-14
Device Manufacturer Date2012-01-01
Date Added to Maude2013-03-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMIKE PHILLIPS, VP
Manufacturer Street48438 MILMONT DRIVE
Manufacturer CityFREMONT CA 945380000
Manufacturer CountryUS
Manufacturer Postal945380000
Manufacturer Phone5102705890
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameALTERG ANTI-GRAVITY TREADMILL
Product CodeBXB
Date Received2013-03-05
Model NumberM320
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALTERG INC
Manufacturer Address48438 MILMONT DRIVE FREMONT CA 94538 US 94538


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-03-05