WHEELCHAIR ACCESSORY AIRPTMBDY18

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-03-15 for WHEELCHAIR ACCESSORY AIRPTMBDY18 manufactured by Invacare Suzhou.

Event Text Entries

[17132782] (b)(4). It was reported by the territory business manager that the airptmbdy18 seat cushion for a power wheelchair was installed over a silver plastic shroud piece with sharp edges accessible to the end user, the technicians, and the caregivers. Additionally, it was reported that upon removing the cushion, inside the silver shroud, there was a metal seat pan, and on the front there was an u shape cutout, which created a second sharp point posing a potential risk for injury. There was no injury reported.
Patient Sequence No: 1, Text Type: D, B5


[17322500] (b)(4). Only one mdr report will be submitted for this event, although two different complaints were created because of different parts needed to repair the unit, and the file numbers are the following: (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004493922-2013-00587
MDR Report Key3006959
Report Source07
Date Received2013-03-15
Date of Report2013-02-12
Date Mfgr Received2013-02-12
Date Added to Maude2013-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactPATRICIA MEDINA
Manufacturer StreetONE INVACARE WAY
Manufacturer CityELYRIA OH 44035
Manufacturer CountryUS
Manufacturer Postal44035
Manufacturer Phone8003336900
Manufacturer G1INVACARE SUZHOU
Manufacturer StreetUNIT 28 ZONE,200 SUHONG RD SUZHOU INDUSTRIAL PARK
Manufacturer CityJIANGSU, P.R.C. 215021
Manufacturer CountryCH
Manufacturer Postal Code215021
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWHEELCHAIR ACCESSORY
Generic Name890.3910
Product CodeINE
Date Received2013-03-15
Model NumberAIRPTMBDY18
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINVACARE SUZHOU
Manufacturer AddressUNIT 28 ZONE,200 SUHONG RD SUZHOU INDUSTRIAL PARK JIANGSU, P.R.C. 215021 CH 215021


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-03-15