[3383327] A customer reported to beckman coulter (bec) that the unicel dxh slidemaker stainer coulter cellular analysis system leaked approximately 3 ml of clear fluid during shutdown. The leak was not contained within the instrument. The operator was wearing gloves and a lab coat at the time of the event. There was no injury or exposure to open wounds or mucous membranes, and the operator did not seek medical attention. Msds was not reviewed, but the facility has an exposure control plan. No erroneous results were generated.
Patient Sequence No: 1, Text Type: D, B5


[10632743] Bec field service engineer (fse) observed that the source of the leak was the wash collar which needed alignment. The fse performed the wash collar alignment followed by daily checks as well as shutdowns. No further evidence of leaking was observed. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1061932-2013-00342
MDR Report Key3008064
Report Source05,06
Date Received2013-03-18
Date of Report2013-02-17
Date of Event2013-02-17
Date Mfgr Received2013-02-17
Device Manufacturer Date2012-02-01
Date Added to Maude2013-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street11800 SW 147TH AVE.
Manufacturer CityMIAMI FL 33196
Manufacturer CountryUS
Manufacturer Postal Code33196
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNICEL? DXH SLIDEMAKER STAINER COULTER? CELLULAR ANALYSIS SYSTEM
Generic NameSPINNER, SLIDE, AUTOMATED
Product CodeGKJ
Date Received2013-03-18
Catalog Number775222
Lot NumberNA
ID NumberSOFTWARE V. 1.5.1.0
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2013-03-18

© 2024 FDA.report