CLINICAL CHEMISTRY PHENOBARBITAL 01E08-21

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-04-01 for CLINICAL CHEMISTRY PHENOBARBITAL 01E08-21 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[3322190] The customer stated phenobarbital quality controls were higher than expected on an archtiect c4000 analyzer when reagent lot 85773un12 was in use. The customer began using phenobarbital reagent lot 96734un13 to resolve the issue.
Patient Sequence No: 1, Text Type: D, B5


[10589484] (b)(4). The investigation determined three lots of phenobarbital reagent (52803un12, 62299un12, and 85773un12) have exhibited increased imprecision associated with the aging of the reagents. The issue is driven by flocculation in the r2 reagent. Customers have been instructed to discontinue use of these three lots and destroy any remaining inventory. A replacement lot with reduced dating is available to customers.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2013-00075
MDR Report Key3029405
Report Source05
Date Received2013-04-01
Date of Report2013-02-01
Date Mfgr Received2013-03-23
Device Manufacturer Date2012-07-31
Date Added to Maude2013-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone847937-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction Number1628664-11/19/12-001-R
Event Type3
Type of Report3

Device Details

Brand NameCLINICAL CHEMISTRY PHENOBARBITAL
Product CodeJJC
Date Received2013-04-01
Catalog Number01E08-21
Lot Number85773UN12
Device Expiration Date2013-02-28
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC
Manufacturer Address1921 HURD DRIVE IRVING TX 75038 US 75038


Patients

Patient NumberTreatmentOutcomeDate
10 2013-04-01

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.