BREMER 3D CROWN FULL ASSEMBLY 292301001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,07 report with the FDA on 2013-04-02 for BREMER 3D CROWN FULL ASSEMBLY 292301001 manufactured by Depuy Synthes Spine.

Event Text Entries

[3255637] International affiliate reports that during application of the 3d halo crown assembly, one skull pin penetrated the patient's skull. The pin was removed and replaced and the halo crown was applied. The patient is reported to have not have any health problems and no additional treatment was required. It was reported that the patient's bone quality is not very good.
Patient Sequence No: 1, Text Type: D, B5


[10609320] No lot number or sample was available. Without a lot number, a review of manufacturing records cannot be completed. Without a product sample, we are unable to confirm the reported issue or identify the root cause. Review of complaint data found no emerging trends. No other complaints of this nature have been reported for the past twelve months. The patient is reported to have poor bone quality which may have contributed to the event. At this time, the complaint is considered to be closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1526439-2013-14720
MDR Report Key3030249
Report Source01,05,07
Date Received2013-04-02
Date of Report2013-03-05
Date of Event2013-03-01
Date Mfgr Received2013-03-05
Date Added to Maude2013-04-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactBETH BECOTTE
Manufacturer Street326 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal02767
Manufacturer Phone5088283575
Manufacturer G1DEPUY SYNTHES SPINE
Manufacturer Street325 PARAMOUNT DRIVE
Manufacturer CityRAYNHAM MA 02767
Manufacturer CountryUS
Manufacturer Postal Code02767
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBREMER 3D CROWN FULL ASSEMBLY
Generic NameCOMPONENT, TRACTION, INVASIVE
Product CodeJEC
Date Received2013-04-02
Catalog Number292301001
Lot NumberUNK
OperatorPHYSICIAN
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES SPINE
Manufacturer Address325 PARAMOUNT DRIVE RAYNHAM MA 02767 US 02767


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2013-04-02