UNKNOWN ZIMMER DRILL BIT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,08 report with the FDA on 2013-03-21 for UNKNOWN ZIMMER DRILL BIT manufactured by Zimmer, Inc..

Event Text Entries

[3201421] It was reported that the tip of the drill bit broke off in the patient's patella.
Patient Sequence No: 1, Text Type: D, B5


[10596181] Evaluation summary: the potential field age of the device could not be determined. The manner of use of the drill bit that led to this even tis unknown. Based on the information provided, a definitive cause for this event cannot be determined with certainty. Returned for review is an unknown drill bit that has fractured just below the start of the cutting flutes. Dimensional analysis could not be performed and review of the device history records was not possible as the product and/or lot numbers required for retrieval were unavailable. It is not suspected that the product failed to meet specifications. The investigation could not verify or identify any evidence of product contribution to the reported problem. Based on the available information, the need for corrective action is not indicated. Should additional substantive information be received, the complaint will be reopened. Zimmer, inc. Considers the investigation closed.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1822565-2013-00511
MDR Report Key3034761
Report Source05,06,08
Date Received2013-03-21
Date of Report2013-02-22
Date of Event2013-02-12
Date Mfgr Received2013-02-22
Date Added to Maude2013-05-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKEVIN ESCAPULE
Manufacturer StreetP.O. BOX 708
Manufacturer CityWARSAW IN 465810708
Manufacturer CountryUS
Manufacturer Postal465810708
Manufacturer Phone8006136131
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameUNKNOWN ZIMMER DRILL BIT
Generic NameKNEE INSTRUMENT
Product CodeHTT
Date Received2013-03-21
Returned To Mfg2013-02-25
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerZIMMER, INC.
Manufacturer AddressP.O. BOX 708 WARSAW IN 46581070 US 46581 0708


Patients

Patient NumberTreatmentOutcomeDate
10 2013-03-21

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.