ADVIA CENTAUR XP CORTISOL (COR) ASSAY N/A 04610138

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-04-04 for ADVIA CENTAUR XP CORTISOL (COR) ASSAY N/A 04610138 manufactured by Siemens Healthcare Diagnostics, Inc..

Event Text Entries

[20234742] Discordant advia centaur xp cortisol results were obtained for two samples from the same patient. The samples were tested 30 minutes apart. The results were questioned by the phyician. Repeat testing was performed on both samples. One result was similar to the initial result and one result was higher than the initial result. It is unknown if patient treatment was prescribed or altered. There was no report of adverse health consequences due to the discordant cortisol results.
Patient Sequence No: 1, Text Type: D, B5


[20422321] A siemens field service engineer (fse) was sent to the customer site and performed a full system inspection. The system was fully functional. No conclusion can be drawn. The cause for the discordant cortisol result is unknown. The instrument is performing within specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1219913-2013-00075
MDR Report Key3036128
Report Source05,06
Date Received2013-04-04
Date of Report2013-03-14
Date of Event2013-03-12
Date Mfgr Received2013-03-14
Date Added to Maude2013-06-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. EIMAN SULIEMAN
Manufacturer Street333 CONEY STREET
Manufacturer CityE. WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal02032
Manufacturer Phone5086604603
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer Street333 CONEY STREET
Manufacturer CityE. WALPOLE MA 02032
Manufacturer CountryUS
Manufacturer Postal Code02032
Single Use3
Remedial ActionIN
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADVIA CENTAUR XP CORTISOL (COR) ASSAY
Generic NameCORTISOL IMMUNOASSAY
Product CodeJFT
Date Received2013-04-04
Model NumberN/A
Catalog Number04610138
Lot Number260
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS, INC.
Manufacturer AddressTARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2013-04-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.