TRANSMOTION MEDICAL, INC TMM5-XTB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-03-15 for TRANSMOTION MEDICAL, INC TMM5-XTB manufactured by Transmotion Medical Inc..

Event Text Entries

[3394677] No injury reported. Submitting based on potential. (b)(6) of (b)(6) (surgery center) e-mailed tmm's sales rep stating that they were getting ready for an eye case, and the "tower" (column) started moving on its own. Her concern was that it could have been a safety issue had it occurred during the procedure.
Patient Sequence No: 1, Text Type: D, B5


[10672189] The remedial action was to replace the pendant, which had been damaged due to an unk cause. The e-mail that was sent to tmm stated they also had an issue with the pendants dragging on the ground, which would not have helped. Transmotion medical, inc is designing a new pendant to launch to improve overall durability/quality and to protect the internal printed circuit board/electronics. This solution is currently undergoing etl testing and will be in place in 2013 for production and service components.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2013-00002
MDR Report Key3050539
Report Source06
Date Received2013-03-15
Date of Report2013-03-13
Date of Event2013-02-14
Date Mfgr Received2013-02-13
Date Added to Maude2013-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL, INC
Generic NameTMM5 MOBILE SURGICAL STRETCHER-CHAIR
Product CodeGBB
Date Received2013-03-15
Model NumberTMM5-XTB
Catalog NumberTMM5-XTB
Lot Number4718
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-03-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.