Event Text Entries

[3252634] Panel personal computer fell off mount, hanging from cables. There was no patient or user injury, but if they were below the computer when it fell, they could have been hurt.
Patient Sequence No: 1, Text Type: D, B5


[10666451] When product is shipped to the customer, there are several sets of screws and instructions explaining which screws to use where. The customer inserted incorrect shorter screws which caused the pc to fall. We have added clearer installation instructions starting in (b)(6) 2013 to systems shipped.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3008289288-2013-00003
MDR Report Key3060308
Report Source05
Date Received2013-04-11
Date of Report2013-04-09
Date of Event2013-03-22
Date Mfgr Received2013-03-22
Device Manufacturer Date2012-12-01
Date Added to Maude2013-04-17
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTONI GORMAN
Manufacturer Street1850 DEMING WAY
Manufacturer CityMIDDLETON WI 53562
Manufacturer CountryUS
Manufacturer Postal53562
Manufacturer Phone6088298600
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNICOLET ONE MONITOR ON ERGO CART
Generic NameNICOLET ONE MONITOR ON ERGO CART
Product CodeOLV
Date Received2013-04-11
Model Number982A0500
Catalog NumberNA
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS NEUROLOGY INCORPORATED
Manufacturer Address1850 DEMING WAY MIDDLETON WI 53562 US 53562


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-04-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.