IBEAM P10105-101

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2006-07-26 for IBEAM P10105-101 manufactured by Medical Intelligence Medizintechnik-gmbh.

Event Text Entries

[17126176] The pt was in prostate treatment and was treated with a brainlab novalis system directly with skin on the couch top. The kind of treatment was conformal beam. During the treatment, the couch top "caused that the patient got an incorrect dose treatment and has dry desquamation". Region of treatment: prostate. Applied radiation energy and modality of 6 mv. Six field technique was used and field size was roughly 6 cm x 7 cm. Delivered dose per fraction of 2 gy at 90%.
Patient Sequence No: 1, Text Type: D, B5


[17323636] The medical intelligence couch top is like, e. G. , siemens, sinmed, and q-fix, a solid carbon fibre couch top without a tennis racket area. (b)(4). All carbon fiber tabletops result in an increase in pt skin dose and all are aware of the increase in dose. This design enables a higher stability for these couch tops and minimizes any sagging of the pt, which can occur with traditional tennis racket inserts or foil and a notably improvement imaging quality, compared to tennis racket types. Over the complete treatment area, solid carbon fiber couch tops do not have a higher dose attenuation compared to tennis racket areas. As a consequence of higher attenuation caused by the solid carbon fibre couch tops, there is a, in comparison, notably higher skin dose. Even very thin carbon fiber solid inserts, to be placed in the tennis racket area, have an notable effect on skin dose up to (b)(4) relative surface dose. Same can be noted with several state of the art immobilization devices. Corresponding (b)(4) calculations demonstrate that this effect to skin dose can be decisively compensated only by tennis rackets. Therefore, all solid carbon fibre couch tops can be stated to show similar effects. The effect has to be considered for adaption of treatment techniques. Should the user employ the same techniques, as for a couch top with tennis racket, skin reactions are more likely to occur. Opposite, couch tops with a tennis racket area have massive frame structures, which have to be considered in treatment planning as well. This imposes that couchtops may not be treated "as air".
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9710332-2006-00004
MDR Report Key3068309
Report Source08
Date Received2006-07-26
Date of Report2006-07-20
Date of Event2006-06-28
Date Mfgr Received2006-06-28
Device Manufacturer Date2005-11-01
Date Added to Maude2013-04-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetFEYERABENDSTRASSE 13-15
Manufacturer CitySCHWABMUNCHEN 86830
Manufacturer CountryGM
Manufacturer Postal86830
Manufacturer Phone232969220
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIBEAM
Product CodeJAI
Date Received2006-07-26
Model NumberNA
Catalog NumberP10105-101
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMEDICAL INTELLIGENCE MEDIZINTECHNIK-GMBH
Manufacturer AddressSCHWABMUNCHEN 86830 GM 86830


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-07-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.